Know-how Ther­mo­cou­ples

Ther­mo­cou­ples are among the most robust, ver­sa­ti­le, and wide­ly used sen­sors in indus­tri­al tem­pe­ra­tu­re mea­su­re­ment tech­no­lo­gy. Whe­ther deal­ing with extre­me heat in fur­nace con­s­truc­tion or deman­ding pro­cess moni­to­ring – the right sel­ec­tion deter­mi­nes mea­su­re­ment accu­ra­cy and lon­ge­vi­ty.

In this sec­tion, we pro­vi­de you with in-depth exper­ti­se: from the phy­si­cal fun­da­men­tals and the See­beck effect to sel­ec­ting the appro­pria­te ther­mo­cou­ple types and moun­ting methods for your appli­ca­ti­on.

Table of Con­tents: Ther­mo­cou­ples

What is a ther­mo­cou­ple?

A ther­mo­cou­ple con­sists of two dif­fe­rent metal­lic con­duc­tors joi­n­ed tog­e­ther at a mea­su­ring point. This junc­tion of two metals is refer­red to as a ther­mo­cou­ple pair (or ther­mo­cou­ple junc­tion).

Ope­ra­ting Prin­ci­ple & Mea­su­ring Prin­ci­ple

  • The See­beck effect: When the ther­mo­cou­ple is expo­sed to a tem­pe­ra­tu­re gra­di­ent, the ther­mo­elec­tric effect gene­ra­tes a mini­mal elec­tri­cal vol­ta­ge (ther­mo­elec­tric vol­ta­ge in the mV or µV ran­ge).
  • Tem­pe­ra­tu­re dif­fe­rence: In addi­ti­on to the mea­su­ring junc­tion, the sys­tem has a refe­rence junc­tion (cold junc­tion) at the other end. The mea­su­red ther­mo­elec­tric vol­ta­ge essen­ti­al­ly depends on the tem­pe­ra­tu­re dif­fe­rence bet­ween the mea­su­ring junc­tion and the refe­rence junc­tion.
  • Mate­ri­al pro­per­ties: The sel­ec­ted com­bi­na­ti­on of con­duc­tor mate­ri­als decisi­ve­ly deter­mi­nes:
    – the magni­tu­de of the gene­ra­ted ther­mo­elec­tric vol­ta­ge
    – the spe­ci­fic ther­mo­elec­tric pro­per­ties
    – the usable tem­pe­ra­tu­re mea­su­re­ment ran­ge

The gene­ra­ti­on of ther­mo­elec­tric vol­ta­ges is based on the See­beck effect: a tem­pe­ra­tu­re gra­di­ent along a metal­lic con­duc­tor gene­ra­tes a local elec­tro­mo­ti­ve force (EMF). In this pro­cess, elec­trons pre­fe­ren­ti­al­ly dif­fu­se from the warm to the cold regi­on, resul­ting in a char­ge sepa­ra­ti­on. The cold end beco­mes nega­tively char­ged, while the warm end beco­mes posi­tively char­ged. The resul­ting elec­tric field oppo­ses fur­ther elec­tron dif­fu­si­on until a ste­ady-sta­te equi­li­bri­um is estab­lished. The EMF cor­re­sponds to the poten­ti­al dif­fe­rence pre­sent in this sta­te of equi­li­bri­um.

In a sim­pli­fied, clas­si­cal model, this elec­tron dif­fu­si­on can be explai­ned by the hig­her avera­ge “kine­tic ener­gy” of the elec­trons in the warm regi­on. Con­se­quent­ly, more elec­trons dif­fu­se from the warm to the cold end than in rever­se.

From a quan­tum mecha­ni­cal per­spec­ti­ve, the occu­pa­ti­on of elec­tro­nic sta­tes dif­fers bet­ween the warm and cold regi­ons. At the cold end, fewer sta­tes are occu­p­ied abo­ve the Fer­mi ener­gy than at the warm end. As a result, more unoc­cu­p­ied sta­tes are available into which elec­trons from the warm regi­on can tran­si­ti­on than vice ver­sa, lea­ding to char­ge sepa­ra­ti­on.

For a con­duc­tor with a tem­pe­ra­tu­re-depen­dent See­beck coef­fi­ci­ent S(T), the EMF gene­ra­ted bet­ween the tem­pe­ra­tures Tcold and Twarm can be for­mal­ly descri­bed by the inte­gral:

GUE_KnowHow_TE_Seebeck_Formel_01_EN

In a clo­sed cir­cuit made of a sin­gle homo­ge­neous mate­ri­al, all EMF con­tri­bu­ti­ons can­cel each other out, resul­ting in no mea­sura­ble total vol­ta­ge. If, howe­ver, two dif­fe­rent con­duc­tors (mate­ri­al A and mate­ri­al B) are con­nec­ted to each other, their respec­ti­ve EMF con­tri­bu­ti­ons sub­tract.

Bet­ween the open ends, a resul­ting ther­mo­elec­tric vol­ta­ge UTh is pro­du­ced:

GUE_KnowHow_TE_Seebeck_Formel_02_EN

This inte­gral illus­tra­tes the under­ly­ing phy­si­cal rela­ti­onship, but is not sui­ta­ble as a prac­ti­cal equa­ti­on for tem­pe­ra­tu­re deter­mi­na­ti­on, as S(T) is non-line­ar and lacks a clo­sed-form phy­si­cal descrip­ti­on.

Negle­c­ting the tem­pe­ra­tu­re depen­dence of SA and SB for small tem­pe­ra­tu­re dif­fe­ren­ces yields the appro­xi­ma­ti­on for UTh:

GUE_KnowHow_TE_Seebeck_Formel_03_EN

For pre­cise tem­pe­ra­tu­re mea­su­re­ments, howe­ver, this sim­pli­fi­ca­ti­on is ina­de­qua­te, as the actu­al tem­pe­ra­tu­re depen­dence of S(T) can­not be igno­red.

In prac­ti­ce, tem­pe­ra­tu­re is the­r­e­fo­re not deter­mi­ned using the See­beck coef­fi­ci­ents direct­ly, but via the refe­rence func­tions spe­ci­fied in DIN EN 60584-1, which mathe­ma­ti­cal­ly link the ther­mo­elec­tric vol­ta­ge direct­ly to the mea­su­red tem­pe­ra­tu­re (see Tem­pe­ra­tu­re Deter­mi­na­ti­on and Refe­rence Func­tions).

Ther­mo­cou­ple Types and Color Coding

The­re are various inter­na­tio­nal stan­dards defi­ning the most com­mon ther­mo­cou­ple types and their color coding. The most pro­mi­nent include DIN EN 60584-3 (Euro­pean), DIN 43714 (Ger­man, obsolete/withdrawn), and ASTM E230/E230M (US). They spe­ci­fy a color code for each ther­mo­cou­ple type to ensu­re unam­bi­guous iden­ti­fi­ca­ti­on and assign­ment of sen­sors, exten­si­on cables, and con­nec­tion com­pon­ents.
Type Ther­mo­cou­ple DIN EN 60584-3 DIN 43714

Type R

Pt13Rh-Pt

DIN EN 60584 Type R and S

DIN 43714 Type R S

Type S

Pt10Rh-Pt

DIN EN 60584 Type R and S

DIN 43714 Type R S

Type B

Pt30Rh-Pt6Rh

DIN EN 60584 Typ B

Type J

Fe-CuNi

DIN EN 60584 Typ J

Type T

Cu-CuNi

DIN EN 60584 Typ T

Type E

NiCr-CuNi

DIN EN 60584 Typ E

Type K

NiCr-Ni

DIN EN 60584 Typ K

DIN 43714 Type K

Type N

NiCr­Si-NiSi

DIN EN 60584 Typ N

Type C

W5Re-W26­Re

DIN EN 60584 Typ C

Type A

W5Re-W20­Re

DIN EN 60584 Typ A

Type L

Fe-CuNi

DIN 43714 Type L

Type U

Cu-CuNi

DIN 43714 Type U

Type Ther­mo­cou­ple ASTM E230/E230M
Ther­mo­cou­ple Cables
ASTM E230/E230M
Com­pen­sa­ting Cables

Type R

Pt13Rh-Pt

ASTM E230 TL Type R S

Type S

Pt10Rh-Pt

ASTM E230 TL Type R S

Type B

Pt30Rh-Pt6Rh

ASTM E230 TL Type B

Type J

Fe-CuNi

ASTM E230 TL Type J

ASTM E230 AGL Type J

Type T

Cu-CuNi

ASTM E230 TL Type T

ASTM E230 AGL Type T

Type E

NiCr-CuNi

ASTM E230 TL Type E

ASTM E230 AGL Type E

Type K

NiCr-Ni

ASTM E230 TL Type K

ASTM E230 AGL Type K

Type N

NiCr­Si-NiSi

ASTM E230 TL Type N

ASTM E230 AGL Type N

Type C

W5Re-W26­Re

ASTM E230 TL Type C

Pro­per­ties and Typi­cal Indus­tri­al Sec­tors

The fol­lo­wing over­view shows the most com­mon­ly used ther­mo­cou­ple types S, B, K, and N along with their respec­ti­ve tem­pe­ra­tu­re ran­ges accor­ding to DIN EN 60584, advan­ta­ges, dis­ad­van­ta­ges, and typi­cal fields of appli­ca­ti­on.
Type Tem­pe­ra­tu­re Ran­ge Advan­ta­ges and Dis­ad­van­ta­ges Typi­cal Indus­tri­al Sec­tors

Type S

0 to 1600 °C

Advan­ta­ges:
Very high mea­su­re­ment accu­ra­cy, low drift, good long-term sta­bi­li­ty

Dis­ad­van­ta­ges:
High cost, low ther­mo­elec­tric vol­ta­ge, sus­cep­ti­ble to con­ta­mi­na­ti­on if not ade­qua­te­ly pro­tec­ted

Type B

600 to 1700 °C

Advan­ta­ges:
Very high long-term sta­bi­li­ty at high tem­pe­ra­tures, low drift

Dis­ad­van­ta­ges:
High cost, very low ther­mo­elec­tric vol­ta­ge, not sui­ta­ble for low tem­pe­ra­tures, sus­cep­ti­ble to con­ta­mi­na­ti­on if not ade­qua­te­ly pro­tec­ted

Type K

-200 to 1200 °C

Advan­ta­ges:
Cost-effec­ti­ve, wide tem­pe­ra­tu­re ran­ge, wide­ly used, ver­sa­ti­le appli­ca­ti­on

Dis­ad­van­ta­ges:
K-effect: drift in the medi­um tem­pe­ra­tu­re ran­ge

Type N

-200 to 1200 °C

Advan­ta­ges:
Fur­ther deve­lo­p­ment of Type K: lower drift, wide tem­pe­ra­tu­re ran­ge, ver­sa­ti­le appli­ca­ti­on

Dis­ad­van­ta­ges:
Less wide­ly used than Type K and slight­ly more expen­si­ve (~15% more expen­si­ve)

In addi­ti­on, the­re are fur­ther ther­mo­cou­ple types – R, J, T, E, C, and A – which are sui­ta­ble for spe­cia­li­zed appli­ca­ti­ons or less com­mon ope­ra­ting are­as.

Type Tem­pe­ra­tu­re Ran­ge Advan­ta­ges and Dis­ad­van­ta­ges Typi­cal Indus­tri­al Sec­tors

Type R

0 to 1600 °C

Advan­ta­ges:
Very high long-term sta­bi­li­ty, low drift, high mea­su­re­ment accu­ra­cy

Dis­ad­van­ta­ges:
High cost, sus­cep­ti­ble to con­ta­mi­na­ti­on if not ade­qua­te­ly pro­tec­ted

Type J

-40 to 750 °C

Advan­ta­ges:
High ther­mo­elec­tric vol­ta­ge, cost-effec­ti­ve

Dis­ad­van­ta­ges:
Not sui­ta­ble for very high tem­pe­ra­tures

Type T

-200 to 350 °C

Advan­ta­ges:
High mea­su­re­ment accu­ra­cy at low tem­pe­ra­tures

Dis­ad­van­ta­ges:
Not sui­ta­ble for high tem­pe­ra­tures

Type E

-40 to 900 °C

Advan­ta­ges:
Very high ther­mo­elec­tric vol­ta­ge

Dis­ad­van­ta­ges:
Drift at high tem­pe­ra­tures, K-effect: drift in the medi­um tem­pe­ra­tu­re ran­ge

Type C

450 to 2300 °C

Advan­ta­ges:
Sui­ta­ble for extre­me­ly high tem­pe­ra­tures

Dis­ad­van­ta­ges:
Only usable under inert gas (pro­tec­ti­ve atmo­sphe­re) or vacu­um

Type A

450 to 2300 °C

Advan­ta­ges:
Sui­ta­ble for extre­me­ly high tem­pe­ra­tures

Dis­ad­van­ta­ges:
Only usable under inert gas (pro­tec­ti­ve atmo­sphe­re) or vacu­um

Basic Con­s­truc­tion

An over­view of the various designs and con­fi­gu­ra­ti­ons of ther­mo­cou­ples can be found in our pro­duct over­view.
A pro­duct fin­der is also available the­re to assist you in sel­ec­ting a sui­ta­ble sen­sor.

Tem­pe­ra­tu­re Deter­mi­na­ti­on and Refe­rence Func­tions

In Euro­pe, tem­pe­ra­tu­re deter­mi­na­ti­on using ther­mo­cou­ples is gene­ral­ly per­for­med based on DIN EN 60584-1. For each ther­mo­cou­ple type, this stan­dard defi­nes refe­rence func­tions UTh(T) that mathe­ma­ti­cal­ly link the ther­mo­elec­tric vol­ta­ge UTh to the tem­pe­ra­tu­re T. The­se refe­rence func­tions are descri­bed by n-th order poly­no­mi­als and repre­sent the non-line­ar rela­ti­onship bet­ween ther­mo­elec­tric vol­ta­ge and tem­pe­ra­tu­re: the poly­no­mi­al coef­fi­ci­ents ai are nor­ma­tively defi­ned for each ther­mo­cou­ple type. Based on the­se refe­rence func­tions – or their inver­se func­tions – the cor­re­spon­ding mea­su­re­ment tem­pe­ra­tu­re can be deter­mi­ned from the mea­su­red ther­mo­elec­tric vol­ta­ge:Formel Temperaturbestimmung The poly­no­mi­al coef­fi­ci­ents ai are nor­ma­tively spe­ci­fied for each ther­mo­cou­ple type. Based on the­se refe­rence func­tions – or their inver­se func­tions – the cor­re­spon­ding mea­su­re­ment tem­pe­ra­tu­re can be deter­mi­ned from the mea­su­red ther­mo­elec­tric vol­ta­ge.

Reference Functions Curve Plot

Tole­rance Clas­ses and Limit Devia­ti­ons

Per­mis­si­ble limit devia­ti­ons (tole­ran­ces) for ther­mo­cou­ples are spe­ci­fied in DIN EN 60584-1 and divi­ded into three accu­ra­cy clas­ses: Class 1 (hig­hest accu­ra­cy), Class 2, and Class 3. GÜNTHER GmbH Tem­pe­ra­tur­mess­tech­nik stri­ves to always sup­p­ly the best pos­si­ble accu­ra­cy class for the respec­ti­ve tem­pe­ra­tu­re ran­ge. In excep­tio­nal cases—for exam­p­le, with noble mine­ral-insu­la­ted thermocouples—only Class 2 or Class 3 can be sup­pli­ed. Reasons for this include mecha­ni­cal stress and che­mi­cal con­ta­mi­na­ti­on, which can affect the ther­mo­elec­tric pro­per­ties.

Fur­ther­mo­re, addi­tio­nal stan­dards app­ly in spe­ci­fic indus­tries, such as the auto­mo­ti­ve indus­try (CQI-9) or aero­space (AMS2750), which defi­ne par­ti­al­ly stric­ter requi­re­ments.

Note: A ther­mo­cou­ple mee­ting Class 1 requi­re­ments does not neces­s­a­ri­ly ful­fill Class 3 requi­re­ments, as the clas­si­fi­ca­ti­ons cover dif­fe­rent tem­pe­ra­tu­re ran­ges.

Type Class 1 Class 2 Class 3

Type R

Tem­pe­ra­tu­re ran­ge:
0 to 1600 °C

Limit devia­ti­on:
T < 1100 °C: 1,0 °C
T > 1100 °C:
1,0 °C + 0,3 % · (T – 1100 °C)

Tem­pe­ra­tu­re ran­ge:
0 to 1600 °C

Limit devia­ti­on:
1,5 °C or 0,25 % · T

Type S

Tem­pe­ra­tu­re ran­ge:
0 to 1600 °C

Limit devia­ti­on:
T < 1100 °C: 1,0 °C
T > 1100 °C:
1,0 °C + 0,3 % · (T – 1100 °C)

Tem­pe­ra­tu­re ran­ge:
0 to 1600 °C

Limit devia­ti­on:
1,5 °C or 0,25 % · T

Type B

Tem­pe­ra­tu­re ran­ge:
600 to 1700 °C

Limit devia­ti­on:
1,5 °C or 0,25 % · T

Tem­pe­ra­tu­re ran­ge:
600 to 1700 °C

Limit devia­ti­on:
4,0 K or 0,5 % · T °C

Type J

Tem­pe­ra­tu­re ran­ge:
-40 to 750 °C

Limit devia­ti­on:
1,5 °C or 0,4 % · T

Tem­pe­ra­tu­re ran­ge:
-40 to 750 °C

Limit devia­ti­on:
2,5 °C or 0,75 % · T

Type T

Tem­pe­ra­tu­re ran­ge:
-40 to 350 °C

Limit devia­ti­on:
0,5 °C or 0,4 % · T

Tem­pe­ra­tu­re ran­ge:
-40 to 350 °C

Limit devia­ti­on:
1,0 °C or 0,75 % · T

Tem­pe­ra­tu­re ran­ge:
-200 to 40 °C

Limit devia­ti­on:
1,0 K or 1,5 % · T

Type E

Tem­pe­ra­tu­re ran­ge:
-40 to 800 °C

Limit devia­ti­on:
1,5 °C or 0,4 % · T

Tem­pe­ra­tu­re ran­ge:
-40 to 900 °C

Limit devia­ti­on:
2,5 °C or 0,75 % · T

Tem­pe­ra­tu­re ran­ge:
-200 to 40 °C

Limit devia­ti­on:
2,5 °C oder 1,5 % · T

Type K

Tem­pe­ra­tu­re ran­ge:
-40 to 1000°C

Limit devia­ti­on:
1,5 °C or 0,4 % · T

Tem­pe­ra­tu­re ran­ge:
-40 to 1200 °C

Limit devia­ti­on:
2,5 °C or 0,75 % · T

Tem­pe­ra­tu­re ran­ge:
-200 to 40 °C

Limit devia­ti­on:
2,5 °C or 1,5 % · T

Type N

Tem­pe­ra­tu­re ran­ge:
-40 to 1000 °C

Limit devia­ti­on:
1,5 °C or 0,4 % · T

Tem­pe­ra­tu­re ran­ge:
-40 to 1200 °C

Limit devia­ti­on:
2,5 °C or 0,75 % · T

Tem­pe­ra­tu­re ran­ge:
-200 to 40 °C

Limit devia­ti­on:
2,5 °C or 1,5 % · T

Type C

Tem­pe­ra­tu­re ran­ge:
426 to 2315 °C

Limit devia­ti­on:
1,0 % · T

Type A

Tem­pe­ra­tu­re ran­ge:
1000 to 2500 °C

Limit devia­ti­on:
1,0 % · T

For tole­ran­ces with “or” spe­ci­fi­ca­ti­ons, the grea­ter value always appli­es.

Typi­cal Sources of Error

When mea­su­ring tem­pe­ra­tu­re with ther­mo­cou­ples, various sources of error can lead to dis­tor­ted or unre­lia­ble mea­su­re­ment results. The most important cau­ses are sum­ma­ri­zed below:

Rever­sed Pola­ri­ty of Con­nec­tors or Wiring
If a plug is con­nec­ted back­wards to its mating socket or if the cable is mis­wired, the thermocouple’s pola­ri­ty beco­mes rever­sed. This can lead to sub­stan­ti­al mea­su­re­ment errors, typi­cal­ly mani­fest­ing as a nega­ti­ve tem­pe­ra­tu­re rea­ding.

Incor­rect Type of Con­nec­tor or Cable
If ther­mo­cou­ple con­nec­tors, exten­si­on cables, or com­pen­sa­ting cables of a dif­fe­rent type than the ther­mo­cou­ple in use are employ­ed, the resul­ting ther­mo­elec­tric vol­ta­ge chan­ges due to dif­fe­rent alloy pro­per­ties, lea­ding to mea­su­re­ment devia­ti­ons.

Incor­rect Type Sel­ec­ted in the Mea­su­ring Device
Indi­vi­du­al ther­mo­cou­ple types have dif­fe­rent refe­rence func­tions cor­re­la­ting ther­mo­elec­tric vol­ta­ge to tem­pe­ra­tu­re. If the instru­ment is set to a type that does not match the con­nec­ted ther­mo­cou­ple, incor­rect tem­pe­ra­tu­re values will be cal­cu­la­ted.

Inho­mo­ge­neous or Con­ta­mi­na­ted Ther­mo­cou­ple Mate­ri­al
If the ther­mo­cou­ple wire mate­ri­al is inho­mo­ge­neous or con­ta­mi­na­ted by for­eign sub­s­tances, this local­ly alters the ther­mo­elec­tric pro­per­ties. This gene­ra­tes para­si­tic ther­mo­elec­tric vol­ta­ges, which can result in dis­tor­ted tem­pe­ra­tu­re rea­dings.

Aging Effects
Over exten­ded peri­ods of use – espe­ci­al­ly at high tem­pe­ra­tures – the mate­ri­al pro­per­ties of the ther­mo­cou­ple wires can chan­ge per­ma­nent­ly and irrever­si­bly, for exam­p­le due to struc­tu­ral chan­ges in the micros­truc­tu­re or dif­fu­si­on pro­ces­ses. This affects the ther­mo­elec­tric pro­per­ties, lea­ding to a gra­du­al drift and mea­su­re­ment inac­cu­ra­ci­es.

K-Effect (Short-Ran­ge Orde­ring)
Type K ther­mo­cou­ples can exhi­bit the so-cal­led “K-effect” within a tem­pe­ra­tu­re ran­ge of appro­xi­m­ate­ly 250 °C to 600 °C. This phe­no­me­non stems from micros­truc­tu­ral chan­ges in the NiCr leg that can occur during ther­mal cycling within this ran­ge and are par­ti­al­ly rever­si­ble. The­se chan­ges alter the ther­mo­elec­tric pro­per­ties, which can result in cycle-depen­dent mea­su­re­ment errors.

Elec­tro­ma­gne­tic Inter­fe­rence
Ther­mo­cou­ples gene­ra­te very small signal vol­ta­ges that are sus­cep­ti­ble to elec­tro­ma­gne­tic fields. Adja­cent elec­tri­cal com­pon­ents can indu­ce addi­tio­nal vol­ta­ges, ther­eby caus­ing cor­rupt­ed tem­pe­ra­tu­re values.

Poor Ther­mal Cou­pling
Insuf­fi­ci­ent ther­mal cou­pling bet­ween the tem­pe­ra­tu­re sen­sor and the medi­um to be mea­su­red impairs heat trans­fer. This can lead to slug­gish mea­su­re­ment respon­se times or ste­ady-sta­te mea­su­re­ment errors.